- Home
- »
- Products
- »
- Impedance Analyzers
- »
- Zeus
- »
- Performance Demonstrations
Performance Demonstrations of Zeus
In the figures below, we present measurement outcomes obtained through the Zeus Electrochemical Impedance Analyzer with FR-EX3, Enhanced Limitation and HI-PR extended options, All measurements are done with 50cm BNC Cables between sample holder and analyzer. Sampling time 0.5s or 1 Cycle.
Zeus – short circuit calibration standard – measured impedance <1mΩ in Basic Range
Zeus – 100Ω calibration standard – measured phase ~1m° in Basic Range
Zeus – open electrodes – measured capacity is around 1fF in Basic Range
Zeus – 100pF calibration standard – measured Tan δ < 1e-4 in Basic Range – Correctly compensated in high frequency
Zeus – 1nF Polyester Capacitor – undercompensated with an inductance of a wire in series
Zeus – 330µH inductor – the model calculates to about 15pF of parallel capacitance inline
Zeus – 10 kΩ Resistor
Zeus – 100 GΩ Resistor
Zeus – Resistor Capacitor network ([100pF + 10Ω] || [470pF || 1.65MΩ ])
Zeus – Resistor Capacitor network (100pF || [470pF + 20kΩ] || [2nF + 10MΩ])