Performance Demonstrations of Omega – HV500 Interface

In the figures below, we present measurement outcomes obtained through Omega Advanced Modular Frequency Analyzer with HV500 High Voltage Impedance Dielectric Interface 2 Wire Front Connection with FR-EX3.

All measurements are done with: 
• 100cm BNC Cables between analyzer and Interface
• 50cm SHV Cables between sample holder and analyzer
• Sampling time 0.5s or 1 Cycle.
• Auto Referencing mode enabled

Omega – HV500 – short circuit calibration standard – measured impedance < 50mΩ in Basic Range

Omega – HV500 – open SHV connector – measured capacity < ~1fF in Basic Range

Omega – HV500 – 100Ω calibration standard – measured phase < ~10m° in Basic Range

Omega – HV500 – 100pF calibration standard – measured Tan δ < 6e-5 in Basic Range – Correctly compensated in high frequency

Omega – HV500 – 1nF Polyester Capacitor – overcompensated in the high frequency range

Omega – HV500 – 10µF Electrolytic Samwha 100V Capacitor – It is measured from high frequency with 50V applied bias with an imposed 35Vrms which equates to an applied Voltage that swings from 0 to 100Vpp to test the real limits of the capacitor

Omega – HV500 – 1µF tantalum Capacitor – Unkown Voltage So measured with a 15V applied bias and 10Vrms until it suffered a breakdown

Omega – HV500 – Resistor Capacitor network (100pF || [470pF + 20kΩ] || [2nF + 10MΩ])