Performance Demonstrations of Omega – HV1000 Interface

In the figures below, we present measurement outcomes obtained through Omega Advanced Modular Frequency Analyzer with HV1000 High Voltage Impedance Dielectric Interface 2 Wire Front Connection with FR-EX3.

All measurements are done with: 
• 100cm BNC Cables between analyzer and Interface
• 50cm SHV Cables between sample holder and analyzer
• Sampling time 0.5s or 1 Cycle.
• Auto Referencing mode always enabled

Omega – HV1000 – short circuit calibration standard – Measured impedance < 0.5Ω

Omega – HV1000 – open SHV connector – measured capacity < ~10fF in Basic Range

Omega – HV1000 – 100Ω calibration standard – measured deviation in impedance <= 0.5Ω

Omega – HV1000 – 100pF Ceramic Capacitor – measured Tan δ < 1e-4 in Basic Range

Omega – HV1000 – 1nF Polyester Capacitor

Omega – HV1000 – 330pF Polyester Capacitor – measured with +490 Vdc and 5Vrms applied voltage

Omega – HV1000 – SMAJ54CA – Bidirectional ESD Protection Diodes – Measures within it’s specs according to a breakdown Voltage of 60 V

Omega – HV1000 – Resistor Capacitor network (1nF || [470pF + 20kΩ] || [2nF + 10MΩ])

Omega – HV1000 – 100pF calibration standard – measured with 10 mVrms applied voltage