In the figures below, we present measurement outcomes obtained through Omega Advanced Modular Frequency Analyzer with HV1000 High Voltage Impedance Dielectric Interface 2 Wire Front Connection with FR-EX3.
All measurements are done with:
• 100cm BNC Cables between analyzer and Interface
• 50cm SHV Cables between sample holder and analyzer
• Sampling time 0.5s or 1 Cycle.
• Auto Referencing mode enabled
Omega – HV1000 – short circuit calibration standard – Measured impedance < 2Ω
Omega – HV1000 – open SHV connector – measured capacity < ~1fF in Basic Range
Omega – HV1000 – 100Ω calibration standard – measured deviation in impedance < 0.5Ω in Basic Range
Omega – HV1000 – 100pF calibration standard – measured Tan δ < 8e-5 in Basic Range
Omega – HV1000 – 1nF Polyester Capacitor
Omega – HV1000 – 470pF Polyester Capacitor – measured with +491 Vdc and 5Vrms applied voltage
Omega – HV1000 – 1µF tantalum Capacitor – Unknown Voltage So measured until it suffered a breakdown at 10 Vrms
Omega – HV1000 – 2x 1N4148 against each other. Worked as a Zener diode in both polarities at 100 Vrms
Omega – HV1000 – Resistor Capacitor network (100pF || [470pF + 20kΩ] || [2nF + 10MΩ])
Omega – HV1000 – 100pF calibration standard – measured with 10 mVrms applied voltage
Omega – HV1000 – 100pF calibration standard – measured with 1 mVrms applied voltage – Still identifiable until low frequencies when SNR is very low